Abstract

To evaluate the relationship between radiation doses and lung computed tomographic (CT) fluoroscopic scan parameters and to determine optimal scan parameters for performance of lung interventional radiologic (IR) procedures. The institutional review board approved this prospective study, which included 32 patients with a single lung tumor; written informed consent was obtained. CT fluoroscopic images were obtained with three tube voltages (80,120,135 kV) and three tube currents (10, 20, 30 mA) in each patient. The signal-to-noise ratios (SNRs) and the contrast-to-noise ratios (CNRs) were measured quantitatively. To evaluate the feasibility of performing lung IR procedures, four readers visually scored the image quality. Acceptable CT fluoroscopic images were determined by using agreement of at least three of the four readers. The weighted CT dose index for each CT scan parameter was measured. A piecewise linear regression equation was obtained from the relationship between radiation doses and visual image scores. Both the SNR and the CNR improved as the radiation dose increased, leading to improvement in the image quality. Acceptable image quality was achieved in 94% (30 of 32) of patients when the radiation dose was 1.18 mGy/sec (120 kV, 10 mA) and in all patients when it was greater than 1.48 mGy/sec (135 kV, 10 mA). The piecewise linear curve showed rapid improvement in image quality until the radiation dose increased to 1.48 mGy/sec (135 kV, 10 mA). When the radiation dose was increased greater than 1.48 mGy/sec, improvement in the image quality became more gradual. Results of this study can be used to guide the determination of optimal scan parameters in lung CT fluoroscopy.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call