Abstract

Locations of surface electromyography (sEMG) electrodes in the face are usually chosen on a macro-anatomical basis. In this study we describe optimal placement of bipolar electrodes based on a novel method and present results for lower facial muscles. We performed high-density sEMG recordings in 13 healthy participants. Raw sEMG signals were decomposed into motor unit action potentials (MUAPs). We positioned virtual electrode pairs in the interpolated monopolar MUAPs at different positions along muscle fiber direction and calculated the bipolar potentials. Electrode sites were determined where maximal bipolar amplitude was achieved and were validated. Objective guidelines for sEMG electrode placement improve the signal-to-noise ratio and may contribute to reduce cross talk, which is particularly important in the face. The method may be regarded as an important basis for improving the validity and reproducibility of sEMG in complex muscle areas.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.