Abstract

When a biocortical screw is used for the rigid fixation of sagittal split-ramus osteotomy, by definition the threads of the screw should engage in the buccal plates as well as in the lingual plates. In a study of six dry mandibles we found that when 2 and 2.7 mm screws are used in noncountersunk holes, it is almost certain that a thread will engage the buccal cortical plate. With countersinking 55% of the 2.7 mm and 27% of the 2.0 mm screws had no threads engaged in the buccal cortical plate. This changes these screws from bicortical to compression type. An anatomic study of bone thickness demonstrated that the thickest buccal and lingual cortical plates were in the superior border of the ramus, just distal to the last molar. In addition, a change in the intercondylar distance before and after surgery was observed in every mandible.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.