Abstract

The photon energies which maximize the signal to noise ratio per unit absorbed dose (optimal photon energies) have been calculated for detection of soft tissue masses within adipose tissue 3--15 cm thick. The detection task simulates mammography in women with fatty breasts. Ideal, plain film and calcium tungstate screen-film systems were considered in the analysis. Net photographic density is not incorporated into this model; therefore, the optimal photon energy with respect to the signal to noise ratio per unit absorbed dose may not be the same as the photon energy which results in maximum film density per unit absorbed dose. Therefore, the model has limited applicability to film or screen-film systems. However, it is directly applicable to digital systems where brightness and contrast can be adjusted.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.