Abstract

This study was undertaken to determine the optimum number of Doppler velocity waveforms required to calculate cardiac output in atrial fibrillation with the same degree of accuracy as that for sinus rhythm. Twenty-one patients in atrial fibrillation underwent calculations of cardiac output derived from aortic Doppler waveform velocity time integrals and RR intervals. The variability in estimates of the cardiac output was calculated with the successive addition of sequential beats and compared with that determined in a control group of 12 subjects in sinus rhythm. For the group in atrial fibrillation, a mean of 13 beats (range 4 to 17 beats) was required to achieve an estimation of cardiac output with a variability of less than 2%, compared with a mean of four beats in sinus rhythm. In atrial fibrillation, the mean number of beats required to determine cardiac output was approximately three times that necessary in sinus rhythm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.