Abstract

Cardiovascular disease is the leading cause of death in the world and the number one killer in Indonesia, with a mortality rate of 17.05%. The target of this research is to increase the range of electrocardiograph (ECG) equipment using LoRa Technology. With LoRa Technology, it is expected that the data transmission process can run effectively and produce an accurate ECG signal and minimal noise. The research method is by sending a heart signal from the ECG simulator by the microcontroller via LoRa Technology which is received by the PC (Personal Computer) and the ECG signal is displayed on the PC display. The most optimal setting will be obtained from the sender-receiver distance and baudrate by measuring data loss and delay. In this study, the simulated cardiac signal from the phantom ECG is fed to an analog signal processing circuit, then the signal is converted to digital and digitally filtered on the microcontroller, then the signal is sent via the LoRa HC-12 Transceiver to a PC with baudrate, distance and barrier settings. The results obtained are that data transmission can be carried out at a distance of 175 meters without a barrier and a distance of 50 meters with a barrier. This remote ECG equipment can detect heart signals and the results can be sent to a PC using LoRa Technology. The implication is that the transmission of ECG signal data via the Lora HC-12 Transceiver media can be carried out optimally at the 9600 baudrate setting.