Abstract

Surface plasmon resonance biosensors measure the thickness or molecular concentration of a biolayer by analyzing small changes in measured reflection spectra. In this paper, we describe linear spectral analysis techniques designed to produce measurements with the maximum possible signal-to-noise ratio. We show how, under appropriate assumptions, an optimal analysis method may be derived for measuring any system parameter, and how measurements of multiple parameters may be made independent in exchange for an decrease in signal to noise ratio. Compared to two conventional data analysis techniques (quadratic fit and centroid methods) using simulated data, the linear techniques show a 30% increase in signal to noise ratio. In application to actual thiol binding data, the linear method yields a signal to noise ratio 46% greater than that of the centroid method and 65% greater than that of the quadratic fit method. This level of noise reduction was achieved by using the ability of the linear methods to reject noise caused by light source brightness variations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.