Abstract

This study verifies the feasibility and optimal conditions for using a general-purpose x-ray imaging device with a relatively large detector and excellent image quality compared to dental x-ray imaging devices according to international standards. The dental material test specimens were produced in the form of disks from four companies with different radiopacity, and aluminum step wedges was manufactured according to ISO 13116. In order to find the optimal irradiation conditions for general X-ray imaging devices, the distance between the target and the tube was fixed at 700 mm, and the exposure time was fixed at 100 ms. And the tube voltage was 50, 60, 70 kV and the tube current was changed to 40, 80, and 160 mA. As a result, it was sufficient to use general-purpose X-ray imaging devices, and the best results were obtained under the optimal conditions of 60 kV and 40 mA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.