Abstract

Cadmium-zinc-telluride detectors enable shorter acquisition durations in myocardial perfusion imaging (MPI), but the time interval of sequential scanning is still unchanged in clinical practice. We designed a very rapid 1-day protocol of MPI using cadmium-zinc-telluride single-photon emission tomography and evaluated the optimal dose ratio between two scanning acquisitions by means of simulations and phantom experiments. We intended to perform a 1-day MPI within 140 min and simulate radioactivities in the second scan under various injected dose ratios. To apply this, a cardiac phantom was scanned with various radioactivities and scans were compared with a reference scan with the ideal tracer concentrations. In the stress-first protocol, the dose ratio 1 : 5 was enough to show the same regional percentage uptake compared with the reference. However, in the rest-first protocol, the regional percentage uptakes were higher than those of the reference image even with a 1 : 6 dose ratio. The injected dose ratio 1 : 5 is optimal in a stress-first rapid 1-day protocol. The rest-first protocol is not appropriate because a dose ratio greater than 1 : 6 is required to withdraw shine-through artifacts.