Abstract

In this study, we used a finite element method to evaluate the maximum Von Mises stresses in jaw bones of immediately loaded implant with different thread heights and widths, and the maximum displacements in implant–abutment complex. The implant thread height ranged from 0.20 to 0.60 mm, and the thread width ranged from 0.10 to 0.40 mm. Compared to those in standard designed implants, the maximum Von Mises stresses in cortical and cancellous bones with axially loaded implants decreased by 18.85% and 47.46%, respectively, and by 16.38% and 63.46%, respectively in buccolingually loaded implants. The maximum displacement of implant–abutment complex loaded axially and buccolingually decreased by 13.78% and 6.97%, respectively. These results indicated that thread height played more important roles in affecting bone stresses and implant–abutment complex stability than thread width. Immediately loaded cylinder implants with thread height exceeding 0.44 mm and width ranging from 0.19 to 0.23 mm caused the lowest stresses to the type B/2 bone.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.