Abstract

This study investigated the optimal design of a radio-frequency (RF) bone tumor ablation device to achieve uniform heating. In a previous study, we confirmed the feasibility of this device, which could heat all regions of the bone to 70°C or higher and maintain this temperature for more than 30min. However, the temperature in each part was non-uniform. To address this issue, the shape of the electrode must be modified to create a uniform electric field. The design of the electrode was optimized to reduce temperature deviations. It is difficult to analytically model the relationship between the shape of the electrode and the electric field. The electrode's design factors were fine-tuned using the Taguchi method, a robust design of experiment approach. The primary objective in this optimization was to maximize the signal-to-noise ratio for temperature in each component, aiming for higher values. After four trials, the signal-to-noise ratio increased in comparison with the initial modified shape from 68.3 to 98.6. The experiment was conducted using an experimental device fabricated using the optimal design factors. In comparison to the previous experiment, the temperature standard deviation per part over time decreased from 10.56°C 4.28°C. The experimental results demonstrated the validity of the proposed optimal design approach. In the future, the proposed method can be used to optimize the design factors when a product is advanced to develop a device that can be applied to the human body.