Abstract

Wireless power transfer (WPT) technology has many applications by virtue of its intrinsic safety and convenience characteristics. This is particularly so when the WPT systems are used in conjunction with implanted medical devices, as they can greatly reduce the pain of patients and their financial burdens. In previous researches, the optimal system mainly focuses on the design of the coils which includes the transmitter coil and the receiver coil. In this paper, the WPT system will be optimized to include also the power source circuit and the load circuit, in addition to the coils, as an integral system. A simple circuit which converts the DC source into AC excitation as the feeding circuit is included in the WPT system. In this convert circuit, there are two inductors which can be formed by coils. In the WPT system, the two coils in the convert circuit are transmitter coils. In addition, the receiver coil should be optimized. Thus, the WPT system is constructed by the three coils. In this paper, the parameters in the convert circuit are considered along with the coil parameters in the optimization study.