Abstract
An optimal control system for the intra-aortic balloon pump (IABP) is presented. Control of the IABP is based on a performance index formulated to reflect a tradeoff between maximising cardiac oxygen supply and minimising cardiac oxygen consumption. In the performance index, mean diastolic pressure (MDP) was used to represent oxygen availability and peak systolic pressure (PSP) was used to represent oxygen consumption. An algorithm, implemented using an 8-bit microcomputer, changes the deflation time of the IABP to maximise this performance index by using an optimisation technique that employs both a search and an approximation. The search produces three equally spaced points which define a region that includes the maximum of the performance index. From these points, the optimum deflation time is estimated by a quadratic approximation. The algorithm has been successfully tested using performance index curves generated by computer simulations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Medical & Biological Engineering & Computing
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.