Abstract

Contrast is used in CT-guided epidural steroid injections to ensure proper needle placement. Once injected, undiluted contrast often obscures the needle, hindering subsequent repositioning. The purpose of this investigation was to establish the optimal contrast dilution for CT-guided epidural steroid injections. This investigation consisted of an initial phantom study, followed by a prospective, randomized, single-center trial assessing a range of contrast dilutions. In the phantom study, a phantom housing a chamber containing a 22-gauge needle and various dilutions of contrast was scanned, and images were evaluated for needle visibility. On the basis of these results, concentrations of 66, 100, 133, and 150 mg/mL iodine were selected for evaluation in a clinical study. Patients presenting for CT-guided epidural steroid injections were randomly assigned to a contrast dilution, and images from the procedure were evaluated by 2 readers blinded to the contrast assignment. Needle visibility was scored by use of a 5-point scale. In the phantom study, the needle was not visible at contrast concentrations of ≥133 mg/mL. In the clinical study, needle visibility was strongly associated with contrast concentration (P < .0001). Significant improvements in visibility were found in 66 mg/mL and 100 mg/mL compared with higher iodine concentrations; no difference was found comparing 66 mg/mL with 100 mg/mL iodine. Neither injection location (cervical versus lumbar) nor technique (interlaminar versus transforaminal) influenced visibility scores. For CT-guided epidural steroid injections, the optimal contrast concentration is 66-100 mg/mL iodine. Because these concentrations are not commercially available, proceduralists must dilute their contrast for such procedures.

Highlights

  • BACKGROUND AND PURPOSEContrast is used in CT-guided epidural steroid injections to ensure proper needle placement

  • Significant improvements in visibility were found in 66 mg/mL and 100 mg/mL compared with higher iodine concentrations; no difference was found comparing 66 mg/mL with 100 mg/mL iodine

  • When the pattern of injected contrast shows that the needle position is not ideal, the proceduralist may reposition the needle

Read more

Summary

Methods

This investigation consisted of an initial phantom study, followed by a prospective, randomized, singlecenter trial assessing a range of contrast dilutions. A phantom housing a chamber containing a 22-gauge needle and various dilutions of contrast was scanned, and images were evaluated for needle visibility. Patients presenting for CT-guided epidural steroid injections were randomly assigned to a contrast dilution, and images from the procedure were evaluated by 2 readers blinded to the contrast assignment. Needle visibility was scored by use of a 5-point scale. This investigation was conducted in 2 parts. The second part was a prospective, randomized, single-center trial assessing a range of contrast dilutions in actual CTF-guided epidural steroid injections. Needle visibility was assessed by use of a 3-point scoring system, with a score of 0 indicating clear visibility of the needle against the background contrast, a score of 1 indicating borderline visibility, necessitating adjustment of the

Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.