Abstract

Luminescence excited by a small X-ray generator (Varian VF-50J) instead of the usual 90Sr– 90Y beta source was measured following optical stimulation. The optimal X-ray irradiation conditions have been assessed, including the irradiation dose rate and the thickness of Al-absorber, in the equivalent dose determination when a luminescence dating process is simulated with laboratory dosed quartz samples. The equivalent doses were found to greatly depend on both the thickness of Al-absorber and the X-ray dose rate. We suggest that the optimal X-ray irradiation conditions for accurate luminescence dating work can be obtained by applying the following: (i) use of an Al-absorber with a thickness of 200 μm and (ii) X-ray dose rates of 0.38 Gy/min to 2.9 Gy/min.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.