Abstract

BackgroundThe spectrum dynamics (SD) algorithm is a cardiac reconstruction algorithm of D-SPECT, which improves spatial resolution compared with the ordered-subsets expectation maximization (OSEM) algorithm. We evaluated the wall thickness and left ventricular (LV) volume in patients with hypertrophic cardiomyopathy (HCM) using the SD algorithm. MethodsIn a phantom study, the myocardial wall was scanned with varying wall thicknesses (10-40 mm). In the clinical study, 30 and 27 normal and HCM patients underwent myocardial perfusion imaging. ResultsIn a phantom study, LV volume using the SD algorithm was increased by thickening the wall of the phantom. In the clinical study, the wall thickness and LV volume of OSEM and SD algorithms showed a difference between HCM and normal groups. The wall thickness using OSEM and SD algorithms were 19.4 ± 2.0 and 16.7 ± 1.5 mm in patients with normal, and 27.9 ± 4.9 and 21.8 ± 2.6 mm in patients with HCM. ConclusionThe SD algorithm in cases of HCM may not be able to correctly assess wall thickness and LV volume. Our study suggests that the OSEM is more suitable in cases of HCM than the SD algorithm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.