Abstract

The design and performance of flow cells and flow cell optics are discussed, with special consideration of the requirements of absorption measuring instruments used as detectors in liquid chromatography. The origins of refractive-index sensitivity and other flow-dependent disturbances and the role of the field and aperture stops in minimizing these effects are described. Equations are derived to show the interaction of optical parameters and flow cell dimensions, and conditions for optimum cell designs are presented. It is shown that under certain conditions, the acceptance of a flow cell varies directly with the square of the volume and inversely with the fourth power of the optical path length. The results of some experimental measurements are presented for various flow cell optical systems, and it is shown that sensitivity to changing refractive index of the solvent can be reduced to a small residual associated with reflection at solvent-window interfaces.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.