Abstract

Real-time interferometric monitoring of the dip coating process is applied to the study of properties of flowing liquids. Nonvolatile Newtonian oils are considered, allowing validity of a simple model after the steady state is reached where film physical thickness depends on time as t(-1/2). Measurement of two distinct mineral oil standards, under several withdrawing speeds, resulted in kinematic viscosities of 1.17+/-0.03 and 9.9+/-0.2 S (1S = 1 cm2/s). Agreement of these results with nominal values from the manufacturer suggests that interferometric monitoring of dip coating may become a valuable method for accurate, contactless viscometry of liquid films. Advantages and present limitations are discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.