Abstract There is a need for isolated current sources for use in selected bioimpedance measurement circuits. The requirement for good isolation is particularly important in medical settings because of safety concerns. A new circuit for producing voltage-controlled current is presented. Measurements have been made on a prototype and simulations have been done on a SPICE model. The presented circuit is an H-bridge where the output devices are the output photodiodes of high-linearity optocouplers. Five operational amplifiers, four high linearity optocouplers, and passive components are used. Output current capability is ±35 μA with an output impedance that is more than 1 M Ω. It is possible to achieve bandwidths above 1 MHz for small load impedances. This circuit is well suited for medical applications thanks to the isolation in the optocouplers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.