Abstract
Manufacturing complex prosthetic framework on osseointegrated implants requires precision at every step of execution. The purpose of this study was to verify the possibility of applying the technology of image acquisition to determine the spatial position of osseointegrated implants. An optical three-dimensional scanning technique was employed: its measurement systematic error (bias) was calculated by comparing the results with the detection on a coordinates measuring machine. Measurements were carried out on master casts by doing an in vitro simulation of intraoral conditions. This study showed that the bias error value of the three-dimensional optical acquiring system was situated between 14 and 21 microm. As far as the accuracy is concerned, it seems possible to use the three-dimensional image acquisition technology as a valid alternative to traditional impression-making procedures. However, the bias levels obtained in this in vitro study will have to be confirmed in a clinical trial.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.