Abstract

Cavity ringdown (CRD) is employed for the first time, to the best of our knowledge, to precisely measure the optical scattering of highly reflective (HR) optics with measurement sensitivity greatly enhanced via power trapping inside the ringdown cavity. The scattering measurement accuracy is significantly improved by calibrating the photo-detector for the scattering measurement with the low transmittance of the cavity mirror or test HR mirror, which is also accurately measured by CRD. The influence of environmental stray light (such as the probe light scattered by optics and mechanical parts outside the ringdown cavity) and other background noises on the scattering measurement is greatly eliminated by the temporal behavior of the scattering CRD signal. A scattering measurement sensitivity of 4.0 × 10-13 is experimentally achieved with a laser with output power of 12 mW.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.