Abstract

An optical method for measuring the frequency response of pressure transducers is described. The method allows the calculation of the varying pressure of ate pressure. Displacement is measured by a photodetector position sensor with the aid of a laser beam reflected on a mirror attached to the loudspeaker diaphragm. The method is based on the short time response of the optical system and it presents some advantages with regard to the usual dynamic calibration using a microphone reference: (a) it has a good response at low frequencies; (b) it is unaffected by environmental conditions; and (c) the system can easily be recalibrated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.