Abstract

Present status of our developments of x-ray holographic microscopes at SPring-8 BL20XU is described. A combination of the x-ray undulator and a zone plate enabled us to make a coherent x-ray source of around 0.1 μm size. Using this secondary source, two types of x-ray holographic microscopes were investigated. First, a Gabor microscope in point-projection geometry was tested. A tantalum 0.2 μm line-and-space pattern could be resolved. Second, using a zone plate as a beam splitter, a Fourier transform holographic microscope was tested. A tantalum 0.2 μm line-and-space pattern could be observed. Polystyrene beads of 2.8 μmand 0.8 μm in diameter could be observed. In Fourier transform holography, a reconstructed image of a specimen that is located out of the plane of the reference source is blurred. Numerical focusing of such an x-ray hologram could be successfully demonstrated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.