Abstract

The present study was undertaken to investigate the effect of argon plasma treatment on the extent of Foley catheter surface modification. Foley catheter (flexible tube made of silicone rubber, used for urinary catheterization) surface was treated at different argon (Ar) RF plasma powers. The surface modification was assessed by measurement of contact angle. Contact angle study shows decrease in contact angle and increase in surface free energy shows more polar group incorporation at low power plasma treatment. Atomic force microscopy (AFM) revealed an increased average surface roughness proportional to plasma power. Optical emission spectroscopy (OES) shows various states of argon causing surface modification.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.