Abstract

In this paper, we introduce a spectrometer design with crossed double-grating for extracting pure rotational Raman (PRR) lines of N2. The design applies a linear optical fibre array as the entrance slit and a linear array photomultiplier tube (LaPMT) as the detector. The linear dispersion and structural parameters of the proposed configuration were analysed theoretically, and the optical layout of the spectrometer was simulated using the Zemax software. A spectrometer with a 32-channel LaPMT was simulated to obtain the PRR spectrum of N2 when the excitation wavelength was 532 nm. The Stokes and anti-Stokes Raman lines, with rotational quantum states from 0 (2) through 14 (16), were obtained on separate channels of the LaPMT, and the resolution was approximately 0.22 nm per channel.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.