Abstract

Genipin cross-linked gelatin hydrogels with 0.2 M sulphuric acid are radiochromic and have sufficient sensitivity for investigating doses less than 50 Gray. Because of the gelatine cross-links, these gels have sufficient strength to allow removal from the vessel container in which they were cast. Placing the gels in the same liquid that was used for preparing the gel allows the radiochemistry to the same throughout the gels and provides physical support. In this buoyancy neutral environment the gel has the same shape and the preparation vessel. This allows optical CT scanning without wall artefacts due to reflection, refraction and optical activity. A gel was irradiated to dose of 25 Gray with a 10 MV photon beam of 20 × 20 mm cross section. Full 3D optical CT scanning was performed with a Vista 10 optical cone-beam CT scanner. Central beam axis profiles and depth dose agree with diode and parallel plate ion chamber measurements. These results demonstrate that genipin cross-linked gel can be used for accurate 3D dosimetry, including surface dose measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.