An optical-loss measurement system based on a resonant Fabry-Perot cavity at 1.06 microm in vacuum has been developed for independent monitoring of the cavity total loss and the optical absorption loss. Maintenance of cavity resonance over a one-month period allows the assessment of long-term degradation of the cavity optics in the presence of outgassing materials, with sensitivities of 5 ppm/yr for total cavity loss and 2 ppm/yr for cavity absorption loss. Test results for light-emitting diodes, Kapton-insulated cable assemblies, and Vac-seal epoxy adhesive are given. Scaling of these results to the optical performance requirements of LIGO is discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.