Abstract

The optimal therapeutic methods for in-stent restenosis (ISR) after carotid artery stenting (CAS) remains controversial. This study aimed to use optical coherence tomography (OCT) to evaluate the in-stent architectures during endovascular angioplasty/stenting for ISR. Six lesions of ISR after CAS were evaluated by OCT during endovascular angioplasty/stenting. In one lesion, the OCT system could not be crossed because of elongation distal to the ISR lesion. In five lesions, pre-procedural OCT clearly revealed neointimal hyperplasia or neoatherosclerosis. The mean in-stent area stenosis was 84%. After regular balloon angioplasty, tissue compression and dissection of various sizes and layers were detected. After balloon angioplasty (with a mean balloon size of 5.4 mm), the minimum lumen area (from 1.7 ± 0.6 to 11.4 ± 5.3 mm2, p < 0.01) and the minimum in-stent area (12.7 ± 2.6 to 21.8 ± 5.0 mm2, p < 0.01) showed a significant increase. Additional stent was placed in one lesion that developed into a flap by dissection after balloon angioplasty. In another lesion in which sufficient dilatation was not achieved by balloon angioplasty, a major stroke occurred by acute occlusion of the ISR lesion 10 months later. OCT can detect the in-stent architecture of ISR lesions after balloon angioplasty and additional stent placement. However, which dissection should be treated by additional stent remain problematic.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call