A digital volume correlation (DVC)-based optical coherence elastography (OCE) method with inverse compositional Gauss-Newton (IC-GN) algorithm and second-order shape function is presented in this study. The systematic measurement errors of displacement and strain from our OCE method were less than 0.2 voxel and 4 × 10-4, respectively. Second-order shape function could better match complex deformation and decrease speckle rigidity-induced error. Compared to conventional methods, our OCE method could track a larger strain range up to 0.095 and reduce relative error by 30-50%. This OCE method has the potential to become an effective tool in characterising mechanical properties of biological tissue.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.