Abstract

A new optical carbon dioxide sensor based on the colormetric change of pH indicator α-naphtholphthalein with the internal reference fluorescent CIS/ZnS quantum dots (QDs) was developed. The luminescence intensity of CIS/ZnS QDs increased with increasing carbon dioxide concentration. The sensing films consisting of α-naphtholphthalein in polyIBM and CIS/ZnS QDs in polystyrene layer. The sensitivity of optical carbon dioxide sensor is quantified in terms of the ration I 100 /I 0 , where I 100 and I 0 represent the detected luminescence intensities in pure carbon dioxide and pure nitrogen environments, respectively. The experimental result reveals that the new optical carbon dioxide sensor has sensitivity of I 100 /I 0 =13.4.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.