Skin thickness, including the adipose layer, which varies from individual to individual, affects the bone density measurement using light. In this study, we proposed a method to measure skin thickness using light and to correct the bias caused by differences in skin thickness and verified the proposed method by experiments using a phantom. We measured simulated skin of different thicknesses and bovine trabecular bone of different bone mineral densities (BMDs) using an optical system consisting of lasers of 850 and 515nm wavelengths, lenses, and slits. Although the slope of the light intensity distribution formed on the surface of the material when irradiated by the 850nm laser is affected by the thickness of the skin phantom. The difference of the intensity distribution peaks (δy) between the 850 and 515nm lasers was strongly correlated with the thickness of the skin phantom. The coefficient of determination between the measurements and the BMD was improved by correcting the 850nm laser measurements with δy. This result suggests that the method is applicable to optical bone densitometry, which is insensitive to differences in skin thickness.