Abstract

Ultrasonographic optic nerve sheath diameter (ONSD) measurement is an accurate, portable, and non-invasive method of detecting raised intracranial pressure that can also reflect dynamic, real-time changes in intracranial pressure fluctuations. Various studies have shown the mean range of ONSD to vary greatly across different population groups. This study aimed to determine the mean ONSD in healthy Black South African adults. In this cross-sectional study, healthy black South African adult participants underwent optic nerve sheath ultrasound of the right eye, with the diameter being measured at 3 mm behind the retina in two different planes. The average of the two measurements was used to find the mean optic nerve sheath diameter. This measurement was compared to that found in a Canadian adult population, and the effect of age, gender, and co-morbidities on ONSD was assessed. A total of 99 participants were included in this study, of which 39 were male and 60 were female. The mean ONSD was found to be 5.1 mm (SD ± 0.33). This value was significantly higher than the mean ONSD observed in the Canadian population (p < 0.001). There was no significant difference found between the mean ONSD in males and females (p = 0.652), and both age and presence of co-morbidities were not found to significantly correlate with ONSD. (p = 0.693 and p = 0.974, respectively).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call