Abstract
While the European legislation states that laboratories of high-containment must be sealable for fumigation, they do not prescribe a minimal value for airtightness. Starting from a previous study in which we measured the airtightness in 4 BSL-3 laboratories with blower-door tests, we discuss the connection between airtightness and a successful decontamination by fumigation. Biological indicators (BIs) consisting of spores of Geobacillus stearothermophilus on metal disks were laid out in laboratories of different levels of airtightness before performing a fumigation with aerosolized hydrogen peroxide using an automated device, according to the manufacturer's instructions. Incubation of all BI disks placed in the facility with the highest level of airtightness showed complete inactivation of spores. However, in the facility with a lower level of airtightness, not all spores were inactivated. Air leaks might be a factor in the outcome of the decontamination of a room by fumigation, as seen in the laboratory with a lower level of airtightness, but other factors associated with the fumigation process might also be critical for a successful decontamination. We argue that a validation of the decontamination procedure, before first use or after important renovations of a laboratory of high-containment, is a more effective endpoint than reaching a predefined level of airtightness.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Applied biosafety : journal of the American Biological Safety Association
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.