Abstract
The present study was performed to identify the theoretical background for optimal use of the bolus tracking system by analyzing the changes in the initial slope of the aortic time-enhancement curve (TEC). We calculated the contrast medium aortic arrival time (TAR), the time to reach the trigger threshold (effective TAR), the slope of the linear equation of the enhancement unit (enhancement rate), and the time to peak aortic enhancement from the TECs of the pharmacokinetic analysis and retrospective clinical study. In the pharmacokinetic analysis, the enhancement rate-simulated under conditions of injection duration 30 s and iodine load per body weight 500 mg/kg-was 27.1 HU/s. In the clinical study, the enhancement rate was 27.9 +/- 3.0 HU/s. A correlation was found between the TAR and the enhancement rate, indicating that enhancement rates decrease with increasing TAR. It took 22.7 +/- 0.5 s to reach maximum enhancement of the aorta from the trigger threshold of an increase of 100 HU and injection duration at 30 s. We found that cardiac output differences are strongly dependent on the TAR and that most of the differences disappeared during the phase until effective TAR.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.