Abstract
For the first time a segmented n-type HPGe detector was operated directly submerged in liquid nitrogen over a long period. As this kind of detector is envisioned to be used in GERDA phase II, it was operated with a low mass signal cable with snap-contacts and mounted in a low-mass copper holder. The detector performance was stable over 146 days, indicating that neither detector nor contacts deteriorated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.