Abstract

Since the electron-capture detector (ECD) is highly selective, it is imperative to use a more universal ionization detector in conjunction with the ECD in order to detect non-capturing or weakly capturing compounds. Also in an EC study of weakly or moderately strong electron-capturing compounds, it is necessary to identify the EC peak of the compound by identifying the major component with an ionization detector. In this paper we have shown that the pulsed discharge detector can be interchanged between the EC and the helium ionization modes within 4–6 s. The application of this procedure has been illustrated with a mixture of alkane/alkene chlorocompounds. The interchange between EC and argon photoionization modes has also been investigated. The change from EC to argon photoionization also occurs in 4–6 s but the reverse process requires ∼9 s.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.