Abstract

We sought to determine if tympanostomy tube (TT) composition impacts the rate of clearing mucoid plugs. The study used an ex vivo model. TTs with a standard shaft length and inner diameter, varying only by composition materials, were studied. Thirty TTs of each biomaterial (stainless steel, titanium, silicone, fluoroplastic, ion-bombarded silicone, and phosphorylcholine-coated fluoroplastic) were plugged with middle-ear mucus and placed in a model ear chamber. Ofloxacin otic solution was instilled into the chamber to cover the plugged TT. Time to clear each plug was recorded. Ion-bombarded silicone TTs cleared more rapidly than plain silicone TTs (P = 0.0042), but no other statistically significant difference among TT materials was observed. TT composition does not significantly affect the rate or "ease" with which TTs may be opened after they become plugged with mucus ex vivo. The higher rate of plug clearance observed with ion-bombarded silicone relative to untreated silicone suggests that improvements may be possible with alterations in either TT composition or surface preparations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.