Abstract

A novel open-ended coplanar waveguide (CPW) sensor for dielectric permittivity measurement is presented in this paper. The open-ended CPW with flange is designed and fabricated on FR4 substrate. The admittance model that includes the radiation effect was used to describe the open-ended CPW sensor. Accurate dielectric permittivity measurements are obtained by using the three-material calibration technique, air, distilled water, and a 25% ethanol aqueous solution. Validation results on 50% and 75% ethanol aqueous solutions agreed with measurements by the commercially available open-ended coaxial-line probe.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.