Abstract

We configured a long-distance ranging apparatus to test the principle of dual-comb time-of-flight measurement using ultrashort lasers. Emphasis was given to the evaluation of open-air performance quantitatively in terms of the measurement resolution and stability. The test results revealed that our dual-comb asynchronous optical pulse sampling permits micrometer-resolved ranging with a repeatability of 2.05 μm over a 648 m distance in dry weather conditions. Further atmospheric effects were evaluated in three different weather conditions with corresponding Allan deviations. Finally, the capability of simultaneous determination of multiple targets was verified with the potential of advanced industrial applications, such as manufacturing, surveying, metrology, and geodesy.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call