Abstract

The entire fluorescence decay profile during HPLC elution has been directly measured on-the-fly in HPLC at higher sensitivity than in previous literature reports. The fluorescence is excited with the fourth harmonic (266 nm) of a pulsed Nd:YAG laser system and detected broadband with a photomultiplier tube and a digital storage oscilloscope. Detection limits in the range 1-10 ppb are found for several individual polycyclic aromatic hydrocarbons (PAHs) when the total time-integrated fluorescence is analyzed. The chromatograms of PAH mixtures containing 8-10 species were lifetime analyzed with a simple phase plane analysis, in which a single lifetime is determined from the fluorescence decay profile for each point on the chromatogram. The determination of lifetimes under coelution conditions is also illustrated and discussed.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call