Abstract

Two primary explosives involved in groundwater contamination, 2,4,6-trinitrotoluene (TNT) and hexahydro-1,3,5-trinitro-1,3,5-triazine (RDX), were detected on-site at low ppb levels with a semiautomated fiber optic biosensor. Validation of the Analyte 2000 (manufactured by Research International Inc., Woodinville, WA) for TNT and RDX detection was performed at two Superfund sites, Umatilla Army Depot and Naval Surface Weapons Center Crane. Samples from monitoring wells were split for analysis using the fiber optic biosensor on-site and using U.S. EPA SW-846 Method 8330 (reverse-phase high performance liquid chromatography) in an offsite laboratory. The Analyte 2000, a multifiber probe fluorimeter, was coupled to a fluidics unit for semiautomated operation. The fiber optic biosensor assay is based on a competitive fluorescent immunoassay performed on the silica core of a fiber probe. From these studies, the limit of detection was determined to be 5 μg/L for both TNT and RDX. In addition to the field samples...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.