Abstract

The on-line incorporation of cloud point extraction (CPE) to flow injection analysis (FIA) is modified to extract and preconcentrate metal species rapidly, avoiding the formation of hydrophobic complexes, using a mixed micellar medium. Coupling the procedure with chemiluminescence (CL) detection based on the catalytic activity of metal species on the luminol–hydrogen peroxide reaction and enhancing the signal with the presence of a micellar carrier containing bromide ions produces a powerful tool for the preconcentration and determination of metal species at ng l −1 levels. As an analytical demonstration ultratrace concentrations of chromium were conveniently detected and quantified in samples with a complex matrix such as seawater and wastewater. The figures of merit for the determination of chromium were: 0.9–1.6% R.S.D. ( n=5) with detection and quantification limits 0.5 and 2.0 ng l −1, respectively. The calibration graph was rectilinear from 2 to 200 ng l −1 ( r=0.9996, n=6). Several other metal ions were determined in ideal situations, with analogous results.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.