Abstract

A new on-line separation/preconcentration system was developed for the determination of Cu(II) ions by flame atomic absorption spectrometry in urine samples. A newly synthesized chelating resin, by anchoring eriochrome blue black R reagent to Amberlite XAD-16 resin, was used as a packing material for the selective separation/preconcentration of Cu(II) ions. The influence of the parameters on the determination of Cu(II) ions such as pH of sample solution, amount of the resin, eluent type, interfering ions and flow variables was studied. The detection limit of the method was 1.0μgL−1 while precision was 2.3% (n=15) at 50μgL−1 Cu(II) level. The adsorption capacity of the resin was 217μgg−1 Cu(II). The accuracy of the method was proven using TMDA-64 standard lake water and synthetic urine sample. The developed method has been applied successfully to the determination of copper in urine with satisfactory results.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call