Abstract
An on-line separation preconcentration system coupled to electrothermal (graphite furnace) atomic absorption spectrometry was developed. A miniature column packed with iminodiacetic acid ethyl cellulose (IDAEC) was inserted into the loop. A peristaltic pump was used to deliver solutions. A flow of air was driven into the packed column, evacuating it between sample loading, washing and elution. The retained analyte was introduced on-line to graphite furnace using countercurrent elution with HNO 3. The system was applied for the determination of V, Co and Pb in medicinal mineral water samples, and nickel in sea water samples. The detection limits (3σ) were 0.058, 0.022, 0.067, 0.062 μg/l for Co, Pb, V, and Ni, respectively. The R.S.D. ( n=5) was <5% at 0.4–1.0 μg/l concentration range.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.