Abstract
A new chelating sorbent has been developed using Amberlite XAD-2 resin anchored with pyrocatechol through the –CH2– group. This sorbent, characterised by elemental analysis and infrared (IR) spectra, was used as a packing material for a minicolumn in an on-line preconcentration system for cadmium, cobalt and nickel determination. Metal ions were sorbed in the minicolumn, from which it could be eluted directly to the nebulizer-burner system of the flame atomic absorption spectrometer (FAAS). Metals can be eluted from the minicolumn with 0.50 mol L−1 HCl or HNO3. The enrichment factors were 22 (Cd), 23 (Co) and 25 (Ni), for 60 s preconcentration time, and 44 (Cd), 40 (Co) and 48 (Ni), if 180 s preconcentration time was used. Under the optimum conditions, the procedure allowed the determination of cadmium, cobalt and nickel with detection limits of 0.27, 0.59 and 1.29 µg L−1, respectively, when preconcentration periods of 180 s were used. The accuracy of the procedure was sufficient and evaluated by analysing certified reference materials. The method was applied to the analysis of food samples (black tea and rice flour).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.