Abstract
Amperometric biosensors (based on rhodinised carbon electrodes) for glucose, glutamine and glutamate were constructed. The sensors were incorporated into a three cell parallel FIA system and used to monitor the three analytes on-line during two mammalian cell perfusion cultures. All measurements were made simultaneously from undiluted media sample. Use of the FIA system enabled easy and rapid exchange of the sensors, during cultivation. The inclusion of a calibration step, regularly for all sensors, helped to maintain the accuracy of all measurements. Comparison with off-line measurements indicated that all three biosensors operated successfully, providing accurate information.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.