Abstract

The performance of the Ion-Selective Electrode (ISE) for in vitro dissolution testing using biorelevant media was evaluated in this study. In vitro dissolution was carried out using USP apparatus 2 (paddle method) with classical and with updated biorelevant media to simulate the pre- and postprandial states. The ISE was used as an analytical stand-alone system and in combination with a single-point HPLC–UV measurement. A modified method enabling the use of the ISE for very poorly soluble substances is also proposed. In terms of f 2-factor, the results acquired using the ISE for the drug diphenhydramine-HCl were found to be very similar to the results obtained by manual sampling followed by HPLC–UV analysis. In Fed State Simulated Gastric Fluid (FeSSGF), a medium containing 50% milk, the ISE is more practical since the need to separate proteins from the analyte prior to HPLC–UV analysis is eliminated. Further work will be needed to establish ISE methodology for Fed State Simulated Intestinal Fluid (FeSSIF) media. In summary, the ISE has promise as an analytical tool for research and development applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.