Abstract

We report a procedure for the online monitoring of aluminium in drinking water by flow injection analysis. The reaction used is the formation of a complex with morin. Under the working conditions, this can be accomplished in an ethanol-rich hydroalcoholic medium, which modifies the fluorescent characteristics of the complex, allowing the determination of aluminium concentrations higher than 3.1 microgl(-1), with a linear application range between 2 and 250 microgl(-1), an R.S.D. of 2.3% (n=10, 120 microgl(-1)) and a sampling frequency of 90 h(-1). The method can thus be considered one of the most sensitive and fastest for the continuous determination of aluminium. In the presence of anionic surfactants, the sensitivity of the determination is increased. In this form, aluminium is detected at concentrations higher than 2.8 microgl(-1), with a linear application range of 2-50 microgl(-1). The procedure was applied to the analysis of aluminium in drinking, river, and underground water. Under the proposed working conditions, only Fe(III), fluoride and phosphates interfere. The interference of Fe(III) can be avoided with hydroxylamine and that of phosphates and polyphosphates by acid digestion of the samples.