Abstract

A capillary electrophoresis method is proposed to analyze the four most well-known growth hormone-releasing hormone (GHRH) analogs that are misused by athletes. Dimethyl-β-cyclodextrin used as a chiral selector allowed, for the first time, the separation of those basic peptide analogs, including enantiopeptides (sermorelin and CJC-1293) that differ by the chirality of only one amino acid. To increase the method sensitivity, electrokinetic preconcentration methods have been investigated. The large volume sample stacking with polarity switching (PS-LVSS) method with an injected sample volume corresponding to 80% of the capillary one was found superior to the sweeping in terms of signal enhancement factor (SEF). Acid and organic solvent addition to the sample (0.1mM phosphoric acid with 30% methanol) led to a twofold signal improvement, when compared to water as a matrix. We increased capillary dimensions to provide a signal enhancement through the injection of a larger sample volume. Finally, using a combination of the optimized PS-LVSS preconcentration with the chiral capillary zone electrophoresis (CZE), the GHRH analogs were separated and limits of detection between 75 and 200ng/mL were reached. This method was successfully applied to urine after a desalting step. An optimized C18 SPE was used for that purpose in order to provide low sample conductivity (<130µS/cm) and preserve the efficiency of LVSS preconcentration. SEF of 640 was obtained with desalted urine spiked with sermorelin by comparison to the CZE (without preconcentration) method.