Abstract

Accurate measurement of sulfated steroid metabolite concentrations can not only enable the elucidation of the mechanisms regulating steroid metabolism, but also lead to the diagnosis of various related diseases. The present study describes a simple and sensitive method for the simultaneous determination of four sulfated steroid metabolites in saliva, pregnenolone sulfate (PREGS), dehydroepiandrosterone sulfate (DHEAS), cortisol sulfate (CRTS), and 17β-estradiol-3-sulfate (E2S), by online coupling of in-tube solid-phase microextraction (IT-SPME) and stable isotope dilution liquid chromatography–tandem mass spectrometry (LC–MS/MS). These compounds were extracted and concentrated on Supel-Q PLOT capillary tubes by IT-SPME and separated and detected within 6 min by LC–MS/MS using an InertSustain swift C18 column and negative ion mode multiple reaction monitoring systems. These operations were fully automated by an online program. Calibration curves using their stable isotope-labeled internal standards showed good linearity in the range of 0.01–2 ng mL−1 for PREGS, DHEAS, and CRTS and of 0.05–10 ng mL−1 for E2S. The limits of detection (S/N = 3) of PREGS, DHEAS, CRTS, and E2S were 0.59, 0.30, 0.80, and 3.20 pg mL−1, respectively. Moreover, intraday and interday variations were lower than 11.1% (n = 5). The recoveries of these compounds from saliva samples were in the range of 86.6–112.9%. The developed method is highly sensitive and specific and can easily measure sulfated steroid metabolite concentrations in 50 μL saliva samples.